Evaluation of the Protect 200 and Protect 800/900 in a Pulmonology Department

Authors

Novaerus

Facility

Uzsoki Hospital; Budapest, Hungary

Download
Full Report

Objective

To test the efficiency of air dis-infection of the Novaerus devices on bacteria.

Methodology

19 Protect 800 devices and 4 Protect 200 devices were installed in the Pulmonology Department, covering 30 beds in 7 wards and connected high/medium risk areas.

8 Protect 800 devices were installed in the Traumatology, Septic Department, covering 12 beds in 4 wards and connected high risk areas.

T0 background contamination samples were taken before installation of the Novaerus devices. T4 efficiency test samples were taking after 2 months operation of the Novaerus devices.

Summary of Results

Testing reflected an 82% drop in CFU rates and a 93% reduction in fungi count. The air quality now meets the Swiss Class III standard (500 CFU/m3 for general wards).